Vibration Chamber Options

Listed below are the options that are available on CV-Series and AV-Series Vibration Chambers.

Cabinet Options
Rear Sliding Door Rear Sliding door provides access to the back of the chamber.
Chamber Height Adjustment Adjust the height of the chamber to interface with slip table or vertical shaker.
Additional Floor Plug A flexible floor plug allows for the addition interface of the shaker. A solid floor plug allows the chamber to be used a standard temperature cycling and humidity chamber.
Additional Chamber Ports Additional circular, square, or rectangular ports are available in a variety of sizes. Ports may be placed on both sides of the chamber for ease of cable routing.
Glove Ports Two feed through ports, typically placed in chamber door, with protective gloves designed to reach in chamber and turn off/on test devices while chamber is in operation.
Cable Notch If routing cables through side ports is not an option for you, a cable notch may be placed in the front of your chamber.
Additional product shelves are available for each CSZ chamber. Custom shelves and fixtures are also available.
If you prefer a chamber with stainless steel exterior, we can include this as an option on any chamber you request.

Humidity Options
Recirculating Water System For customers that do not have a deionized water supply, we provide a recirculating water system that collects condensate from the chamber, stores and filters it in a reservoir for reuse.
For tests requiring humidity levels below 10% RH, we provide an optional low humidity system for testing down to 5% RH.
Humidity Demineralizer Option Demineralizer removes water impurities and minerals that can harm your test chamber. Cartridges will turn orange-brown when used and are easily replaceable. Additional demineralizer cartridges are available for overnight delivery.
Water Pressure Regulator Water pressure regulator is recommended to regulate incoming water supply to <=25 psig
Reverse Osmosis Water purification system for use with city / tap water. This option is not needed if chamber has demineralizer filter system.

Instrumentation Options
EZ-View Chamber Software CSZ EZ-View software is available for monitoring and controlling up to 20 chambers from any location.
Chamber Hi-Low Limit Provides essential failsafe over/under (hi/low) temperature protection for your valuable test items.
IEEE Interface A parallel interface can be provided for interface with IEEE-488 parallel instrument control buss. A standard IEEE-488 connector is provided for easy connection to adjacent test equipment. This option is cannot be used in conjunction with RS-232/RS-485 communications.
Chart Recorder A selection of one or two pen chart recorders are available with a variety of pre-printed charts or self printing thermal charts.
Run Time Meter The optional run time meter includes a non-resettable display which will accumulate the system running time for routine maintenance and planning purposes.
Main Power Disconnect Provides an additional level of safety by shutting down the main power to the chamber prior to maintenance or repair.
Customer Event Option Includes 15 event outputs with connection panel & removable plug connectors. Designed to turn customer’s electronic DUT’s (devices under test) ON or OFF at various stages of the customer programmed test profile. Each output may also be configured to perform other operations including alarm or profile status indicators.
Digital Input Option Includes connection panel and 8 inputs that can be configured to perform various controller functions (i.e. start, hold, resume or terminate a profile, advance steps, control customer event outputs, etc.)

(provides 8 thermocouple / TC inputs through EZT-570i)

EZT Monitor Option Designed to consist of eight (8) “T” type thermocouples (TC’s) for temperature readout with individual alarms, displays, and data logging for each input that can be monitored by the EZT-570i and the EZ-View software. The “black-box” unit itself will be mounted to the side or rear of the unit and the eight TC’s will pass through one of the existing side ports. These are TC’s only (no RH sensors or PT100-RTD’s) and are used for product temperature monitoring or temperature uniformity check within the chamber. (For monitoring only, no control.)
Watlow F4 Controller CSZ F4 Programmable Controller with Product Monitor, redundant high/low limit and EIA 232/485 Computer Interface. Replaces EZT-570i and mounts on a special fixture below the door.

Performance & Refrigeration Options
LN2 Boost Boost cooling is available for applications that require a fast pulldown rate. An LN2 supply tank is required with a pressure of 25 psi.
CO2/LN2 cooled chambers are available without mechanical refrigeration for lower initial costs on applications requiring ultra low temperatures down to -300°F. An LN2 supply tank is required with a pressure of 25 psi.
Dry Air Purge A dry air purge system can be provided to reduce condensation within the chamber and on the test item during temperature cycling testing. This option is primarily used to prevent short circuits in electronic components and is also used for low humidity requirements.
Allows the testing chamber refrigeration system to be located approximately 50 linear feet away from the chamber. This option is commonly used for labs with space restrictions or to minimize noise levels.
Although CSZ chambers are quiet operating, sound foam is available as an option to further reduce the noise generated by the refrigeration system. This option is commonly used in larger horsepower chambers.